Hach Dual Parameter Panel for Chlorine and pH with SC4500, CL17sc and digital pH Sensor, DPMP-CLPHD

Description
The Hach® Dual Parameter Chlorine panel offers CL17sc chlorine analyzer, Digital pH sensor and SC4500 controller in the convenience of one panel. These panel configurations are pre-engineered, ready to install and simple solutions that are designed to work together.
Features
- One panel for monitoring free or total chlorine residual from 0-10 mg/L and pH with SC4500 controller
- Expertly designed, preassembled panels ease purchasing and startup
- Up and running in minutes - wall or rack mount using (4) bolts, a single 1/4" inlet, a single 3/4" drain and one 115V power cord
- Simplify monitoring and maintenance with streamlined layout
Includes
SC4500 Controller, Claros-enabled, 5x mA Output, 2 digital Sensors, 100-240 VAC, US plug (LXV525.99EA1551), CL17sc Colorimetric Chlorine Analyser with Pressure Regulator Installation Kit, without Reagents (8572400), Digital pH Sensor, PPS, Water Distribution Monitoring Panel Mounting (DPD1R1-WDMP), User manual
Product Specifications
| Display | 3.5-inch TFT color display with capacitive touchpad (SC4500) | 
| Range 1 | Chlorine: 0 -10 mg/L Cl2 | 
| Wavelength | 510 nm | 
| Interferences | Other oxidizing agents such as bromide, chlorine dioxide, permanganate, and ozone | 
| Enclosure Rating | IP66 / NEMA 4X | 
| Accuracy | Chlorine: ± 5% or ± 0.04 mg/L (whichever is greater) from 0 - 5 mg/L Cl₂; ± 10% from 5 - 10 mg/L Cl₂ pH: ± 0.02 pH | 
| Light source | LED, measurement at 510 nm, 1 cm light pathlength | 
| Instrument | CL17sc, DPD1R1, SC4500 | 
| Weight | 40.0000 | 
| Sample Requirements | Y-strainer filtration w/ 40-mesh screen or higher; 100-1000 mL/min; 200-500 (optimal); 4.5-75 psi | 
| Measurement Method | DPD | 
| Power Requirements | 100-240 VAC ±10%, 50/60 Hz; 1 A | 
| Parameters | Chlorine & pH | 
| Lamp Type | LED | 
| Range 2 | pH: -2.0 to 14.0 pH | 