Hach DR6000™ UV VIS Spectrophotometer w/ RFID Technology, LPV441.99.00012
Description
The DR6000™ is the industry’s most advanced laboratory spectrophotometer for water testing. It offers high-speed wavelength scanning across the UV and Visible Spectrum, and comes with over 250 pre-programmed methods including the most common water and environmental testing methods used today. With optional accessories allowing for high-volume testing via a carousel sample changer, and increased accuracy with a sample delivery system that eliminates optical difference errors, this instrument ensures you are ready to handle your wide-ranging water testing needs. Combining the DR6000 with Hach’s patented TNTplus™ test kits, you gain additional accuracy with guided step-by-step testing procedures, while making scratched, flawed or dirty glassware a non-issue due to averaging 10 readings and discarding outliers. The spectrophotometers also utilizes RFID technology, allowing it to let you know automatically if your TNTplus test kits have expired, while detecting coefficient factors to avoid errors which can occur in lot-to-lot variations in the chemistries. Partnering all of these features with an integrated quality assurance software package allows for scheduling, documenting and interpreting your needed quality measurements. Your water and environmental testing needs, all in one spectrophotometer. *RFID currently available only in US, Anguilla, American Samoa, Australia, Bolivia, Canada, Cayman Islands, Columbia, Dominican Republic, El Salvador, Federated States of Micronesia, Guam, Guatemala, Marshall Islands, New Zealand, Northern Mariana Islands, Palau, Panama, Puerto Rico, and US Virgin Islands. Customers in other countries should order LPV441.99.00002.
Features
The DR6000 has the most pre-programmed testing methods, including high-speed wavelength scanning across the UV and Visible Spectrum.
A carousel sample changer allows up to seven sequential measurements.
The DR6000 comes with integrated QA software for scheduling, documenting and interpreting all of your needed quality measurements.
The DR6000, when used with TNTplus® reagent vials, provides the accurate results you need by guiding you step-by-step through your testing procedures.
Product Specifications
Warranty | 12 months |
Display | TFT 7 inch WVGA color touch |
Operating Conditions | 10 - 40 °C, max. 80% relative humidity (non-condensing) |
Wavelength Resolution | 0.1 nm |
Wavelength Reproducibility | 0.1 nm |
Wavelength Selection | Automatic, based on method selection |
Wavelength Accuracy | ± 1 nm |
Wavelength Range | 190 - 1100 nm |
User Programs | 200 |
Cuvette Compatibility | Rectangular: 10, 20, 30, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch. |
Storage Conditions | -25 to 60 °C / max. 80% relative humidity (non-condensing) |
User Interface Languages | bg, cn, cz, da, en, es, fr, gr, hr, hu, it, jp, kr, nl, pl, pt, ro, ru, sl, sl, sv, tr |
Weight | 36.4500 |
Spectral Bandwidth | 2 nm |
Stray Light | KI-solution at 220 nm 3.3 Abs/ 0.05% |
Specific Technology | RFID for easy method update, sample ID and Certificate of Analysis |
Source Lamp | Tungsten (visible range), deuterium (UV range) |
Scanning Speed | 900 nm/min (in 1 nm steps) |
Preprogrammed Methods | 240 |
Photometric Linearity | 0.005 - 2 Abs ≤ 0.01 at 2 Abs with neutral glass at 546 nm |
Photometric Measuring Range | ± 3 Abs |
Photometric Accuracy | 5 mAbs @ 0.0 - 0.5 Abs 1% @ 0.5 - 2.0 Abs @ 546 nm |
Optical System | Reference beam, spectral |
Operating Mode | Transmittance (%), absorbance and concentration (wavelength, time) |