Hach DR 1900 Hach Portable spectrophotometer, DR1900-01H
Accessories
Description
The DR1900 is perfect for use in the field because it is the lightest and most compact portable spectrophotometer. It also can perform that largest number of tests of any portable instrument. Underneath the rugged exterior, the DR1900 has the largest number--over 220--of the most commonly tested preprogrammed methods already built in. You can also use the easy-to-use interface to create your own methods. Also, it is flexible because you can use the widest range of vial sizes of any other spectrophotometer.
Features
- The lightest, most compact portable spectrophotometer available easily goes where you need to go.
- Analyze your water with more preprogrammed methods than any other portable spectrophotometer, or easily create your own methods to suit your unique needs.
- Wavelength range of 340 to 800 nm provides accuracy typically seen only in laboratory instruments.
- Simple, easy-to-use interface plus the widest range of vial sizes makes testing flexible.
- Rugged construction keeps the elements out, which means you can test in the most demanding conditions.
Includes
Includes: portable spectrophotometer, printed basic instrument manual, dust cover, 1-in. square matched glass sample cells, AA alkaline batteries 4/pk, and a four-type adapter set.
Product Specifications
Operating Conditions | 10 - 40 °C (50 - 104 °F), max. 80 % relative humidity (non-condensing) |
Power supply | 4 x AA size Alkaline |
Battery Requirements | 4, AA size alkali cells |
Wavelength Reproducibility | ± 0.1 nm |
Wavelength Selection | Automatic |
Wavelength Accuracy | ± 2 nm (range 340 - 800nm) |
Wavelength Range | 340 - 800 nm |
Wavelength Calibration | Automatic |
User Programs | 50 |
Detector | Silicon photodiode |
Cuvette Compatibility | 13 / 16 mm and 1 inch round adapter, 10 x 10 mm 1 inch square and 10x10mm |
Storage Conditions | -30 to 60 °C (-22 to 140 °F), max. 80 % relative humidity (non-condensing) |
Interface | USB type Mini IP67 (with optional Module) |
Battery Life | 15 days (typical) @ 5 readings a day / 5 day week without backlight |
Stray Light | 0.5 %T at 340 nm with NaNO 2 |
Source Lamp | Xenon Flash |
Reproducibility | ± 0.005 Abs (0 - 1 A) |
Photometric Linearity | 0.5 % (0.5 - 2.0 Abs) |
Photometric Measuring Range | 0 - 3 Abs (wavelength range 340 - 800 nm) |
Photometric Accuracy | ± 0.003 Abs @ 0.0 - 0.5 Abs |
Optical System | Reference beam, spectral |
Operating Mode | Transmittance (%), Absorbance and Concentration |
Temperature range | 0 - 50 °C (32 - 122 °F) |
Warranty | 12 months |