Hach CLF10 sc Free Chlorine Analyzer (Panel Only) Grab Sample, LXV45A.99.11022
Description
The CLF10 sc Free Chlorine Analyzer and CLT10 sc Total Chlorine Analyzer measure free or total chlorine continuously for real time process control. It is based on amperometric technology and is fully compatible with all of Hach's digital controllers. The CLF10 sc can be used in most municipal and industrial applications, and is best utilized where waste stream management is a constraint.
Features
- The CLF10 sc and CLT10 sc chlorine analyzers are compliant with EPA Method 334.0, ensuring that you not only have precise process control, but reportable results as well.
- Hach’s amperometric chlorine technology does not use reagents, eliminating the need for a sanitary drain and saving you the cost of monthly reagent replacement.
- The CLF10 sc and CLT10 sc are plug-and-play with Hach’s digital controllers, ensuring easy setup. Available with exclusive self-diagnostics, these instruments will alert the user to changes in the process or upcoming maintenance. Hach’s Cal Watch feature will provide a warning for pH or chlorine calibration changes, and a non-contacting flow sensor will notify users of insufficient sample flow.
Includes
CL10 sc Panel, 1 m digital extension cable, Panel Manual, Chlorine Sensor Manual.
Product Specifications
Repeatability | 30 ppb or 3%, whichever is greater |
Storage Conditions | -20 °C to 60 °C dry |
Interferences | Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. |
Accuracy | Free Chlorine: ±3% of the reference test (DPD) at constant pH less than 7.2 (±0.2 pH unit) |
Response Time | Free Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH |
Measurement Method | Reagentless, electrochemical, three-electrode amperometric system |
Parameters | Free Chlorine |
Warranty | 12 months |
California residents see Prop 65 WARNINGS